I am attempting to run a batch file with the Shimadzu QP 2010+ and keep receiving the error, "Illegal to call out while inside message filter". What should I do?
Applies to: Chromatography QP2010 GCMS (GC/MS) | Tickets Solved: 0 | Rank: 5.5
Error Message
- This error message indicates that the Cycle Composer and GC/MS Solutions software are not properly communicating.
- The software is indicating that the pathway for the method being used is not correct, making it unable to find the folder.
- This may happen even if the correct pathway is being used.
- In order to correct this error, the operator must change the pathways so that they are incorrect, restart the computer, and then change them back to the correct pathways.
Contact technical support if further assistance is needed
Need more help with this instrument?
Our SMEs are standing by to assist with issues and repairs.
Was this helpful? Yes No
Keywords: batch, method, path, pathway, folder, call out, error