The auto-sampler syringe will not release the sample vials and is unable to inject into the Shimadzu QP 2010+. How can I correct this issue?
Applies to: Chromatography QP2010 GCMS (GC/MS) | Tickets Solved: 0 | Rank: 12.6
Improper Auto-Sampler Function
- The auto-sampler will not release the vials if the magnet for head-space analysis is attached.
- Using the Allen-drive in the Shimadzu tool kit, remove the magnet from the auto-sampler needle guide.
- Once the magnet is removed, conduct a rail check on the auto-sampler by following the step-by-step attached document titled "Auto Sampler Needle Change_5" to ensure the alignment is correct.
Contact technical support if further assistance is needed